Model 2461C and 2251A Command Cross Reference

Racal Instruments has discontinued production of all 2251 and 2251A counters. The new Model 2461C Counter is our closest replacement for this counter. A command cross-reference guide is presented below to compare the command sets of these two counters. The key difference is that the 2461C uses a SCPI command set, while the 2251 and 2251A both use a native command set.

Our experience has seen many smooth transitions between the older 2251 products and the new 2461C. To help with this transition, a specification cross-reference guide has been developed to show how 2251 and 2461C specifications differ.

If you need additional compatibility information, please contact Racal Instruments Product Marketing.

 

2251 Command Equivalent 2461C Command(s) Comments
A0 Default state of 2461C Selects front panel Input A
A1 Not implemented on 2461C Selects SUMBUS Input A signal source
A2 Not implemented on 2461C Connects timebase reference to Input A
AAC INP1:COUP AC Selects Input A AC coupling
AAD INP1:ATT X1 Disables Input A x10 Attenuator
AAE INP1:ATT X10 Enables Input A x10 Attenuator
AAU INP1:COMP:AUTO ON Selects Input A autotrigger mode
ADC INP1:COUP DC Selects Input A DC coupling
AE SENS:AVER ON Enables Averaging mode, 100 samples
AFD INP1:FILT OFF Disables Input A Low Pass Filter
AFE INP1:FILT ON Enables Input A Low Pass Filter
AHI INP1:IMP 1M Sets Input A input impedance to 1Meg ohm
ALI INP1:IMP 50 Sets Input A input impedance to 50 ohms
AMN INP1:COMP:AUTO OFF Selects Input A Manual Trigger Mode
ANS INP1:COMP:SLOP NEG Sets Input A Trigger to Negative Slope
APS INP1:COMP:SLOP POS Sets Input A trigger to Positivie Slope
B0 SOUR:ROSC CLK10 Selects 10 MHz signal from backplane as reference frequency
B1 SOUR:ROSC EXT Selects EXT STD. INPUT as reference frequency
BAC INP2:COUP AC Selects Input B AC coupling
BAD INP2:ATT X1 Disables Input B x10 attenuator
BAE INP2:ATT X10 Enables Input B x10 Attenuator
BAU INP2:COMP:AUTO ON Selects Input B autotrigger mode
BCC INP:ROUT COMM Selects Input A and B "common" mode
BCS INP:ROUT SEP Selects Input A and B "separate" mode
BDC INP2:COUP DC Selects Input B DC coupling
BHI INP2:IMP 1M Sets Input B input impedance to 1Meg ohm
BLI INP2:IMP 50 Sets Input B input impedance to 50 ohms
BMN INP2:COMP:AUTO OFF Selects Input B Manual Trigger Mode
BNS INP2:COMP:SLOP NEG Sets Input B Trigger to Negative Slope
BPS INP2:COMP:SLOP POS Sets Input B trigger to Positivie Slope
CK CONF:CHECk Selects CHECK mode
D0 OUTP:TTLT0 ON Enable Gate out signal on TTLTRG0
D1 OUTP:TTLT1 ON Enable Gate out signal on TTLTRG1
D2 OUTP:TTLT2 ON Enable Gate out signal on TTLTRG2
D3 OUTP:TTLT3 ON Enable Gate out signal on TTLTRG3
D4 OUTP:TTLT4 ON Enable Gate out signal on TTLTRG4
D5 OUTP:TTLT5 ON Enable Gate out signal on TTLTRG5
D6 OUTP:TTLT6 ON Enable Gate out signal on TTLTRG6
D7 OUTP:TTLT7 ON Enable Gate out signal on TTLTRG7
D8 OUTP:TTLT0-7 OFF Disable Gate output (default)
DD TINT:DEL:TIME:STAT OFF Disables Time Interval Delay
DE TINT:DEL:TIME:STAT ON Enables Time Interval Delay
FA CONF1:FREQ Selects Frequency A measurement mode
FB CONF2:FREQ Selects Frequency B measurement mode
FC CONF3:FREQ Selects Frequency C measurement mode
FT CONF1:FTIM Selects Fall Time A measurement mode
IP RES Instrument Preset. Sets counter to power-up default state
MD CALC:STAT OFF Disables Math function (default)
ME CALC:STAT ON Enables Math function
NA SENS:AVER OFF Disables 100 sample averaging mode
NW CONF1:NWID Selects Negative pulse width A measurement mode
PA CONF1:PER Selects Period A measurement mode
PH CONF1:PHAS Selects Phase A relative to B measurement mode
PW CONF1:PWID Selects Positive Pulse Width A measurement mode
Q0 *SRE 0 SRQ assertion unconditionally inhibited
Q1 *ESE 60;*SRE 4 SRQ assertion upon error detection
Q2 *SRE 16 SRQ asserted for measurement ready
Q3 *ESE 60;*SRE 20 SRQ asserted for measurement ready or error detection
Q4 Not implemented in 2461C SRQ asserted for frequency standard change
Q5 Not implemented in 2461C SRQ asserted for frequency standard change or error detection
Q6 Not implemented in 2461C SRQ asserted for frequency standard change or measurement ready
Q7 Not implemented in 2461C SRQ asserted for frequency standard change or error detection or measurement ready
R0 ARM:STAR:SOUR TTLT0 External arm input from TTLTRG0
R1 ARM:STAR:SOUR TTLT1 External arm input from TTLTRG1
R2 ARM:STAR:SOUR TTLT2 External arm input from TTLTRG2
R3 ARM:STAR:SOUR TTLT3 External arm input from TTLTRG3
R4 ARM:STAR:SOUR TTLT4 External arm input from TTLTRG4
R5 ARM:STAR:SOUR TTLT5 External arm input from TTLTRG5
R6 ARM:STAR:SOUR TTLT6 External arm input from TTLTRG6
R7 ARM:STAR:SOUR TTLT7 External arm input from TTLTRG7
R8 ARM:STAR:SOUR EXT Enable external arm input on front panel
RA CONF1:FREQ:RAT Selects Ratio A/B measurement mode
RC CONF3:FREQ:RAT Selects Ratio C/B measurement mode
RDT TINT:DEL:TIME? Recalls arming delay time
RE READ? Resets measurement
RF TOT:GAT:INT Reads total so far
RGS *IDN? (first rev. number listed) Recalls VXIbus Processor firmware revision
RGT SENS:APER? Recalls Gate Time
RLA INP1:COMP:LEV? Recalls Input A trigger level
RLB INP2:COMP:LEV? Recalls Input B trigger level
RMn Not implemented in 2461C Recalls memory setting n. Sets counter to pre-stored configuration.
RMS *IDN? (second rev. number listed) Recalls counter firmware revision
RMX CALC:MATH:EXPR? Recalls math constant X
RMY CALC:MATH:EXPR? Recalls math constant Y
RMZ CALC:MATH:EXPR? Recalls math constant Z
RRS SENS:RES? Recalls measurement resolution
RSF Not needed for 2461C Recalls special function number
RT CONF:RTIM Selects Rise Time A measurement mode
RUT Not implemented in 2461C Recalls unit type
S10 ARM:SOUR INT;STOP:SOUR INT Arming Mode: Start: Internal        Stop: Internal
S11 ARM:SLOP POS;SOUR EXT;STOP:SOUR INT Arming Mode: Start: External +ve Stop: Internal
S12 ARM:SLOP NEG;SOUR EXT;STOP:SOUR INT Arming Mode: Start: External -ve  Stop: Internal
S15 ARM:SLOP POS;SOUR EXT;STOP:SLOP POS;STOP:SOUR EXT Arming Mode: Start: External +ve Stop: External +ve
S16 ARM:SLOP POS;SOUR EXT;STOP:SLOP NEG;STOP:SOUR INT Arming Mode: Start: External +ve Stop: External -ve
S17 ARM:SLOP NEG;SOUR EXT;STOP:SLOP POS;STOP:SOUR EXT Arming Mode: Start: External -ve  Stop: External +ve
S18 ARM:SLOP NEG;SOUR EXT;STOP:SLOP NEG;STOP:SOUR INT Arming Mode: Start: External -ve  Stop: External -ve
S20 RES Selects default configuration of Input A and B/Cancels S21
S21 All meas. Functions are interchangeable Interchanges Inputs A and B
S30 INP:COMP:SET:AUTO:ON Selects continuous auto-trigger measurement
S31 INP:COMP:SET:AUTO:ONCE Selects single shot auto-trigger measurement
S50 CONF1:VOLT:MID Selects Mean Amplitude measurement mode
S51 CONF1:VOLT:MAX Selects Positive Peak amplitude measurement mode
S52 CONF1:VOLT:MIN Selects Negative Peak amplitude measurement mode
S60 SENS1:TOT:GAT ON Selects Totalize A by B measurement mode with TA command
S61 SENS1:TOT:GAT OFF Selects Manual Totalize measurement mode with TA command
S70 CONF:CHECk Enables check mode with CK command
S72 CONF:CHECk;TEST:STAT ON;:TEST: TEC:SHORT ON Start TEC short continuous calibration in CK mode
S73 CONF:CHECk;TEST:STAT ON;:TEST: TEC:LONG ON Start TEC long continuous calibration in CK mode
S74 CONF:CHECk;TEST:STAT ON;:TEST: TEC:SHORT ON Start TEC short calibration in CK mode
S75 CONF:CHECk;TEST:STAT ON;:TEST: TEC:LONG ON Start TEC long calibration in CK mode
S76 CONF:CHECk;TEST:STAT ON;:TEST:DAC ON Check Input A and B DACs (continuously ramped) in CK mode
S77 CONF:CHECk;TEST:STAT ON;:TEST1:RELay ON Input A relay check (requires 10MHz signal) in CK mode
S78 CONF:CHECk;TEST:STAT ON;:TEST2:RELay ON Input B relay check (requires 10MHz signal) in CK mode
S80 Not implemented in 2461C Enables 2 character function header on readback string
S81 Default state of the 2461C Disables 2 character function header on readback string
SDTn TINT:DEL:TIME n Sets arming delay time to n seconds
SFD Not required for 2461C Disables special functions
SFE Not required for 2461C Enables special functions
SGTn APER n Sets gate time to n seconds
SLAv INP1:COMP:LEV v Sets Input A trigger level to v Volts
SLBv INP2:COMP:LEV v Sets Input B trigger level to v Volts
SMn Not implemented in 2461C Stores the present configuration in state n (0 to 9)
SMXa CALC:MATH:EXPR a,b,c Sets math constant X = a
SMYb CALC:MATH:EXPR a,b,c Sets math constant Y = b
SMZc CALC:MATH:EXPR a,b,c Sets math constant Z = c
SRSm RES m Sets measurement resolution in digits (3 to 10)
T0 INIT:CONT ON Selects continuous measurement mode
T1 INIT:CONT OFF Selects Single shot measurement mode
T2 TOT:GAT:MAN Starts Manual Totalize or triggers a measurement cycle in T1 (single shot) mode
T3 TOT:GAT:MAN Stops Manual Totalize
TA CONF:TOT Selects Totalize A by B measurement mode
TI CONF:TINT Selects Time Interval measurement mode
TRG TRIG Triggers the counter to take a measurement
Notes:
2251 Operation 2461C Equivalent Comments
Read output buffer READ? 2251 is constantly updating the input buffer; 2461C does when prompted to
Special functions No action required Nothing special needs to be done to access any 2461C functions

 

 

 

 

 

 

 